How to use ESCR Tester
1- Hot bath2- Pyrex tubes3- Press4- Transfer tool5- Brass channels6- Notch making device7- Bending tool8- Cutter punch for press9- Temperature controller10- Digital timer How to use ESCR tester from AHP
1- Hot bath2- Pyrex tubes3- Press4- Transfer tool5- Brass channels6- Notch making device7- Bending tool8- Cutter punch for press9- Temperature controller10- Digital timer How to use ESCR tester from AHP
How to set SV temperature and PV bias for calibration purposes Set the requested temperature SV as below: How to set input bias for PV display of controller. For this purpose, you need to measure requested temperature with calibrated temperature reading device and set offset value in parameter Inb as below:
4. Summary of Test Method 4.1 Bent specimens of the plastic, each having a controlled imperfection on one surface, are exposed to the action of a surface-active agent. The proportion of the total number of specimens that crack in a given time is observed. 6. Apparatus 6.1 Blanking Die—A rectangular die or other means suitable…
Calibration offset value. Be noted after set of proper value don’t forget to “ENTER”.
Tensile test software has installation folder as below: 1- Click on My Installer folder 2- Click on folder Volume 3- Right click on Setup.Exe and “run as administrator” 4- Then install the software in the default root. Please don’t change the installation default root. 5- Next you need to install the driver for communication and…
Mechanical Tests : There are mechanical tests and physical tests for polyethylene pipes as per mentioned standard as below For above tests that is hydrostatic pressure testing you will need 4 types of equipment: 1- Pressure units (Number of lines are as per customer request and max pressure of testing for PE pipes is less…
Pressure Not Increase in the outlet 1- Check main power key is ON 2- Check software has connection to machine and led is green (refer to hydrostatic software guide post)3- Check sound of electromotor if electric is connected when you start specific line 4- If motor is running when you start a line first of…
Port 1: it is the page related to station number of 1 of the hydro test unit Setpoint1(bar): setpoint of requested pressure in bar Rate1(bar/min): Rate of pressure increase, It is used mainly in burst test of pipes and fittings. because in burst test sample need to be burst in definite time of 60-70s. if…
5 Procedure5.1 Test conditionsCarry out the weightings at standard laboratory temperature (23+-2 C)5.2 Test portionTake three test portions as follows.Weigh, to the nearest 0,000 1 g, approximately 1 g of the material taken from the consignment or from the wall of the pipe or from the fitting, reduced to small fragments.5.3 DeterminationThe determination on each…
1- First download driver from below link 2- Run the file as administrator 3- Extract file 4- Click Next 5- Accept agreement 6- At the end click Finish 7- After installation, refer to (Device Manager/Ports) from control panel of Windows and check if you have virtual COM port 8- All AHP’s computerized products have COM…