Calibration of Hydrostatic Pressure Tester (Pressure and Temperature)
Calibration offset value. Be noted after set of proper value don’t forget to “ENTER”.
Calibration offset value. Be noted after set of proper value don’t forget to “ENTER”.
Pressure Not Increase in the outlet 1- Check main power key is ON 2- Check software has connection to machine and led is green (refer to hydrostatic software guide post)3- Check sound of electromotor if electric is connected when you start specific line 4- If motor is running when you start a line first of…
Port 1: it is the page related to station number of 1 of the hydro test unit Setpoint1(bar): setpoint of requested pressure in bar Rate1(bar/min): Rate of pressure increase, It is used mainly in burst test of pipes and fittings. because in burst test sample need to be burst in definite time of 60-70s. if…
1- Place bath in secure place so that at least 1 meter around it need to be empty for service and repair 2- According to the power of heating of bath you need electric panel with proper circuit breaker small size bath has 6KW power, medium size is 18KW, Big size is 30KW power 3-…
1 Scope This part of ISO 1167 specifies a general test method for determining the resistance to internal hydrostatic pressure at a given temperature of thermoplastics pipes, fittings and piping systems for the transport of fluids. The method accommodates water-in-water, water-in-air and water-in-liquid tests. 4 Principle After conditioning, test pieces are subjected to a specified…
Unpacking of the machines First of all, gently unpack the machine. Normally, a full pack of the machine will come to the customer with: Place the bath and appropriate location in your lab. Be noted hydrostatic pressure testing will have some water vapor from the bath when you open the door and some water outlets…