Installation and Connection to DTCOM Software
Calibration offset value. Be noted after set of proper value don’t forget to “ENTER”. Unless the value will not be saved.
Calibration offset value. Be noted after set of proper value don’t forget to “ENTER”. Unless the value will not be saved.
Port 1: it is the page related to station number of 1 of the hydro test unit Setpoint1(bar): setpoint of requested pressure in bar Rate1(bar/min): Rate of pressure increase, It is used mainly in burst test of pipes and fittings. because in burst test sample need to be burst in definite time of 60-70s. if…
1- Download the installation folder of software 2- Go through below installation file and “Run as administrator” 3- Install the software in default root and don’t change the installation root 4- From start menu Run the program as administrator 4- Then you will see a file name “VICAT.dat” in the root of “C” drive. This…