Installation and Connection to DTCOM Software
Calibration offset value. Be noted after set of proper value don’t forget to “ENTER”. Unless the value will not be saved.
Calibration offset value. Be noted after set of proper value don’t forget to “ENTER”. Unless the value will not be saved.
1- Frist be sure that the driver of machine is installed in the computer [How to install communication driver] 2- Install DTCOM software [How to install DTCOM software]. 3- Open DTCOM software and connect to proper virtual COM port that is installed in No.1 4- Now you have connection to address 1 and 2 as…
Software installation 1- Along with the machine you get a flash disk including software of AHP ROTO 2- In the My Installer there is a folder named “Volume” 3- In this folder just right click “Set up.exe” and select “run as administrator” 4- Don’t change default root for installation and continue till the end 5-…
OIT measurement After installation of driver: And installation of the software: Main panel of software has 4 pages: First of all set the parameters for OIT test as below (got to “parameters Tab”): For OIT measurement you need to set two main parameters of “Setpoint(degC)” and “Rate(degC/min)”. For example for PE OIT measurement temperature is…
Before installation of software you need to install communication driver as below guide. Then you have two folders in the software disk. Don’t forget that flash disk of software is not only the software it is also hardware lock for software and you need to keep it for replacement from company side in even in…
4. Summary of Test Method 4.1 The specimen to be tested and the corresponding reference material are heated from room temperature at a constant rate in a non-purging, high-pressure oxygen environment at a defined pressure. When the specified temperature has been reached, the specimen is then held at that temperature until the oxidative reaction is…
1- Hot bath2- Pyrex tubes3- Press4- Transfer tool5- Brass channels6- Notch making device7- Bending tool8- Cutter punch for press9- Temperature controller10- Digital timer How to use ESCR tester from AHP
How to set SV temperature and PV bias for calibration purposes Set the requested temperature SV as below: How to set input bias for PV display of controller. For this purpose, you need to measure requested temperature with calibrated temperature reading device and set offset value in parameter Inb as below:
4. Summary of Test Method 4.1 Bent specimens of the plastic, each having a controlled imperfection on one surface, are exposed to the action of a surface-active agent. The proportion of the total number of specimens that crack in a given time is observed. 6. Apparatus 6.1 Blanking Die—A rectangular die or other means suitable…
Calibration offset value. Be noted after set of proper value don’t forget to “ENTER”.